Steps for resolution:
- Ensure foaming is not causing a false ‘Low Tank Level’ alarm. Do this by turning the sprayers off, letting sit idle for 1-minute, and checking tank level again.
- Ensure float switch is freely moving. To gain access to the float switch follow steps 1-5 of the Routine Maintenance section.
- If foaming is the cause, follow resolution steps for Defoamer Required
- If foaming is not the cause, in Manual Mode click the DOSE button. When full the machine will stop dosing automatically.
- In the Alarms and Warnings screen hit the CLEAR button
If the issue persists please Request Technical Support