Flexijet’s built-in tilt sensor detects the level of the Flexijet and provides the data for the instant Auto-Levelling command.
As a result, the tripod can be set to close to level, and Flexijet compensates for the fine levelling internally.

Tilt Monitoring also uses the tilt sensor to sense for changes to the level of the Flexijet. This is useful where the surface on which the Flexijet is setup is unstable and has the potential to shift, such as a on scaffolding, or an unstable flooring surface. This function is not usually required for the measuring of countertop surfaces, and can be turned off at the operator’s discretion.

Flexijet will display the Tilt Monitoring window each time it is powered on.

Turning off ‘Tilt-Monitoring’ (GIF)

Tilt Monitoring Warning.

If movement greater than the selected setting is detected, then the laser will display a warning notification.
Tilt Monitoring Warning pop-up on Flexijet screen

To turn Tilt Monitoring off while active:

  1. Tap on the Menu button on the top left corner of the Flexijet’s screen
    Tilt monitoring warning screen
  1. Select ‘Tilt Monitoring’ from the menu list
    Flexijet3D screen showing menu with Tilt Monitoring circled
  1. Turn off ‘Tilt monitoring’ by dragging the small red button to the left, and tap ‘close’ on the bottom of the tilt monitoring window.
    Tilt monitoring with switch circled

Feedback